Detail of Recall

Type of Product:Device
Medsafe Reference:16558
Brand Name:Siemens Digital Linear Accelerator PRIMUS, ONCOR and ARTISTE 550TxT Treatment Table
Model:7346534
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:1. In very rare instances the actual table position maysignificantly deviate from the position at the table side control and the LINAC control console, but no interlock occurs.
2. In sporadic cases the following issue for the table axis lateral, longitudinal and isocentric rotation occurs: After a manual table movement with released brake an immediate automatic movement for the same trable axis is not possible.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory