Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16558 |
| Brand Name: | Siemens Digital Linear Accelerator PRIMUS, ONCOR and ARTISTE 550TxT Treatment Table |
| Model: | 7346534 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | 1. In very rare instances the actual table position maysignificantly deviate from the position at the table side control and the LINAC control console, but no interlock occurs. 2. In sporadic cases the following issue for the table axis lateral, longitudinal and isocentric rotation occurs: After a manual table movement with released brake an immediate automatic movement for the same trable axis is not possible. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/4/2014 |