Detail of Recall

Type of Product:Device
Medsafe Reference:16552
Brand Name:Baxter FOLFusor SV 0.5ml/hr
Model:2C4700K
Affected:all lot numbers up to an including 14B039
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare SA (Switzerland)
Issue:Manufacturer has continued to investigate complaints for over-infusion and wants to make customers aware that the direction insert states that infusion rates may be up to 10% greater than the nominal (labeled) infusion rate of 0.5ml/hr. Recent review of flow rate testing indicates that infusion rates may be higher and could result in a 15% greater than nominal (labeled) infusion rate.
Baxter is considering options to bring the flow rate into alignment with curent labeling.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/4/2014