Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16546 |
| Brand Name: | Siemens ADVIA Centaur Systems TSH3 Ultra 100T, 500T |
| Model: | 06491072 06491080 |
| Affected: | Kit lots ending in 266, 267, 269, 270, 271, 272 and 273 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Brand Name: | Siemens ADVIA Centaur Systems VitD 100T, 500T |
| Model: | 10491994 10699201 10631021 10699533 |
| Affected: | Kit lots ending in: 023 024 025 027 027 050 051 052 053 054 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Brand Name: | Siemens ADVIA Centaur BRAHMS PCT 100T |
| Model: | 10378883 |
| Affected: | Kit lots ending in 020 021 022 023 and 024 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Manufacturer confirmed that samples containing fluoroscein may show inteference with the Advia Centaur systems TSH3 Ultra, Vitamin D and the Advia Centaur BRAHMS Procalcitonin assays. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/4/2014 |