Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16543
Brand Name:Fisher & Paykel Nasal CPAP prongs
Model:BC3020-10, BC3520-10, BC4030-10, BC4540-10, BC5040-10, BC5050-10, BC5550-10, BC5560-10, BC6060-10, BC6070-10, BC6570-10
Affected:multiple
Software version:
Recalling Organisation:Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki
AUCKLAND 2013
Contact Information:0800 503 553
Manufacturer:Fisher & Paykel Healthcare S.A. de C.V.
Brand Name:Fisher & Paykel Bubble CPAP Starter Kits
Model:BC461-SK. BC471-SK, BC490-SK, BC491-SK and BC492-SK
Affected:131007 - 140325
Software version:
Recalling Organisation:Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki
AUCKLAND 2013
Contact Information:0800 503 553
Manufacturer:Fisher & Paykel Healthcare S.A. de C.V.
Issue:Manufacturer has received customer reports relating to certain lot numbers of certain models of the nasal CPAP prongs detaching from the FPH FlexiTrunk Nasal Tubing.
FPH introducred a new version of Nasal CPAP prongs in 2013 and the lot numbers and models incorporate this new version. Since the release of this new version FPH has received reports of the affected prongs detaching from the nasal tubing during use especially in the presence of mucous and/ or moisture.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/4/2014
Hide menus