Detail of Recall

Type of Product:Device
Medsafe Reference:16518
Brand Name:Philips HeartStart XL+ Defibrillator/Monitor
Model:Model number: 861290
Affected:
Software version:Units upgraded to B.00.01 software, or manufactured by Philips between December, 2013 and February, 2014
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:A software communication failure on the HeartStart XL+ that occurs during an automated ready-for-use test or at turn-on may lock the user out of clinical mode, which may delay therapy
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/4/2014