Detail of Recall

Type of Product:Device
Medsafe Reference:16496
Brand Name:Kenex Suspension Arm with Radiation Shield and/or Surgical Lamp
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Kenex Electro Medical Limited
Issue:GE Healthcare has recently become aware of an improper installation of the Kenex suspension arm of a radiation shield. The suspension arm was not installed per specified installation requirements. There has been one reported incident of a radiation shield falling from the ceiling causing a minor injury.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/4/2014