Type of Product:Device
Medsafe Reference:16485
Brand Name:GE Magnetic Resonance Imaging Systems
Model:Models: : Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR750w GEM products
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance can lead to increasing heat during scanning which, in turn, can cause material in the Rear Endbell enclosure to overheat, generate smoke in the scan room, and cause some localized melting of the Endbell. This overheating has occurred only while units were being serviced by GE service engineers during repeated evaluation scanning. However, the issue could also occur in Clinical mode.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2014