Detail of Recall

Type of Product:Device
Medsafe Reference:16467
Brand Name:Siemens ADVIA Centaur Folate
Model:10310308 10325366 10331250 and 10340209
Affected:kit lots ending in 219, 222, 223, 224, 225, 226 and 227
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer has confirmed two different sets of complaints regarding the ADVIA Centaur Systems Folate Asay:
Folate lot 223: Siemens confirmed one set of customer complaints with the ADVIA Centaur Systems Folate assay, kit lots ending in 223. Customers observed that the Bio-Rad Liquichek™ and Bio-Rad Lyphochek™ Immunoassay Plus Folate serum controls may go out of range low with the ADVIA Centaur Systems Folate assay, kit lots ending in 223.
Folate lot 219 Whole blood bias: In addition, Siemens has confirmed a second set of customer complaints regarding a shift in whole blood patient results when moving from ADVIA Centaur Systems Folate assay, kit lots ending in 219 to kit lots ending in 222, 224, 225, 226 and/or 227.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/3/2014