Detail of Recall

Type of Product:Device
Medsafe Reference:16458
Brand Name:Baxter Colleague Volumetric Infusion Pumps
Model:2M91617 and 2M91637
Affected:
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare Corporation SA
Issue:Potential failure of fuses fitted in the rear housing of the Colleague Volumetric Infusion Pump.
During 2011 it was identified that a quantity of pumps fitted with two ceramic 1.6A slow-blow fuses in the rear housing was failing prematurely. Investigation lead to a change of the fuse type fitted during future manufacture and service, from a ceramic fuse to a glass fuse, which has a different rupture characteristic.
The failure rate of this reported issue has subsequently declined following the replacement with a glass fuse.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/3/2014