Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16458 |
| Brand Name: | Baxter Colleague Volumetric Infusion Pumps |
| Model: | 2M91617 and 2M91637 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare Corporation SA |
| Issue: | Potential failure of fuses fitted in the rear housing of the Colleague Volumetric Infusion Pump. During 2011 it was identified that a quantity of pumps fitted with two ceramic 1.6A slow-blow fuses in the rear housing was failing prematurely. Investigation lead to a change of the fuse type fitted during future manufacture and service, from a ceramic fuse to a glass fuse, which has a different rupture characteristic. The failure rate of this reported issue has subsequently declined following the replacement with a glass fuse. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/3/2014 |