Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16451 |
| Brand Name: | Varian Medical Systems On-Board Imager |
| Model: | H08 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Varian Medical Systems Australasia Pty Ltd (NZ) Millenium Centre, Level 3, Building C, 600 Great South Road Ellerslie Auckland 1051 |
| Contact Information: | 0800 486 732 |
| Manufacturer: | Varian Medical Systems Inc |
| Issue: | The shaft of the Type 01 elbow motor for Exact arms used in the On-Board Imager device have failed due to belt over-tensioning. This over tensioning would have taken place during servicing. This failure would allow the Exact arm to swing freely with gravity. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/4/2014 |