Detail of Recall

Type of Product:Device
Medsafe Reference:16451
Brand Name:Varian Medical Systems On-Board Imager
Model:H08
Affected:
Software version:
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Issue:The shaft of the Type 01 elbow motor for Exact arms used in the On-Board Imager device have failed due to belt over-tensioning. This over tensioning would have taken place during servicing. This failure would allow the Exact arm to swing freely with gravity.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/4/2014