Detail of Recall

Type of Product:Device
Medsafe Reference:16450
Brand Name:Stryker Shape Arm Kit Left
Model:0682000976
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Communications Inc
Brand Name:Stryker Shape Arm Kit Right
Model:0682000977
Affected:080909DB001 through 131021DB013
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Communications Inc
Issue:Manufacturer has become aware that there is a remote likelihood of the lock washer and screw that secures the monitor onto the SHAPE Arm not to be present. If not present the monitor could separate and dislodge from the Arm potentially resulting in injury.
The SHAPE Arm is an articulating arm for the Boom system. It is intended for the mounting/ positioning of a flat panel monitor that is used to view/ perform surgery. The monitor is attached to the SHAPE arm by standard VESA/ tilter bracket that is installed onto the SHAPE arm and is secured with a lock washer and screw.
The SHAPE Arm does not sustain or support life.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/3/2014