Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16414
Brand Name:Zoll X-Series Monitor/Defibrillators
Model:
Affected:
Software version:Software Version 02.10.02.00 or higher
Recalling Organisation:Zoll Medical New Zealand
1st Floor Unit 1, 29 Acheron Drive
CHRISTCHURCH
Contact Information:03 341 5016
Manufacturer:Zoll Medical Corporation
Issue:Zoll reports that monitor/defibrillator devices operating software version 02.10.02.00 or higher may enter into a continuous system reset loop preventing the device from being used.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2014