Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16405
Brand Name:Siemens Acuson P300 Ultrasound System with cardiac probe PA230E
Model:Probe model: PA230E
Affected:All users of the P300 + PA230E combination
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:Using the PA230 transducer with a user-defined preset which includes the 100% setting for output power could cause the transducer surface temperature to exceed the maximum allowable limit. The potential risk is a patient burn.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/3/2014