Detail of Recall

Type of Product:Device
Medsafe Reference:16404
Brand Name:Roche cobas c502 c-pack MULTI
Model:C501
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:The manufacturer advises there is potential for erroneous results to be generated by the cobas c502 as a consequence of incorrect cassette filling. Due to an in-house labelling software limitation, the ICVC 'on' settings are not correctly encoded in the REC files released to the market so far, and in consequence no ICVC is being performed on cobas c502 for tests requiring manual reagent handling.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2014