Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16388 |
| Brand Name: | Integra CUSA Excel+ Sterile Torque Wrench, disposable |
| Model: | Catalogue: C5601 |
| Affected: | Lots: 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873, 1135014 |
| Software version: | |
| Recalling Organisation: | Integra Neurosciences Pty Ltd (NZ) 12 Florence Avenue Orewa Auckland 0931 |
| Contact Information: | 0800 400 660 |
| Manufacturer: | Valleylab, a division of Tyco Healthcare Group LP |
| Issue: | There is potential that some lots of 23kHz Sterile Torque Disposable Wrenches contain wrenches that will only fit 36kHz CUSA hand piece tips. The green colour code and size printed on the wrenches are correct for a 23kHz wrench, but the affected wrench size insert is smaller than it should be and will only fit a 36kHz tip. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/3/2014 |