Detail of Recall

Type of Product:Device
Medsafe Reference:16382
Brand Name:Siemens Symbia T and Symbia T2 Camera Systems
Model:
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The manufacturer has identified certain Symbia T and T2 systems that have a known hardware defect inside the CT gantry where the circuit boards may come loose due to fatigue of the mounting rivets. The loose components may result in an internal error causing the CT gantry to be inoperable.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/3/2014