Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16382 |
| Brand Name: | Siemens Symbia T and Symbia T2 Camera Systems |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Medical Solutions USA Inc |
| Issue: | The manufacturer has identified certain Symbia T and T2 systems that have a known hardware defect inside the CT gantry where the circuit boards may come loose due to fatigue of the mounting rivets. The loose components may result in an internal error causing the CT gantry to be inoperable. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/3/2014 |