Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16373 |
| Brand Name: | Siemens Artis Zeego, Connector Robotic Stand |
| Model: | |
| Affected: | serial numbers of the motor connectors lower than 946295 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | Manufacturer confirms that in the course of continuous product monitoring inthe field, it has been discovered that in four individual cases the crimping of motor connectors A1 o A3 was incorrectly carried out. In these four cases incorrectly executed crimping in which the crimping tool was used on the wrong point of the contact led to substandard contacting. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/3/2014 |