Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16373
Brand Name:Siemens Artis Zeego, Connector Robotic Stand
Model:
Affected:serial numbers of the motor connectors lower than 946295
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Manufacturer confirms that in the course of continuous product monitoring inthe field, it has been discovered that in four individual cases the crimping of motor connectors A1 o A3 was incorrectly carried out. In these four cases incorrectly executed crimping in which the crimping tool was used on the wrong point of the contact led to substandard contacting.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/3/2014