Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16363
Brand Name:Medtronic Activa PC Neurostimulator
Model:Model: 37601
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa RC Neurostimulator
Model:Model: 37612
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa SC Neurostimulator
Model:Model: 37602
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa SC Neurostimulator
Model:Model: 37603
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreADVANCED
Model:Model: 37713
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreULTRA
Model:Model: 37712
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreSENSOR
Model:Model: 37714
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic PrimeADVANCED
Model:Model: 37702
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreULTRA SureScan MRI
Model:Model: 97712
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreADVANCED SureScan MRI
Model:Model: 97713
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreSENSOR SureScan MRI
Model:Model: 97714
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic PrimeAdvanced SureScan MRI
Model:Model: 97702
Affected:All devices
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:Medtronic has updated the labelling for Medtronic Deep Brain Stimulation (DBS), Spinal Cord Stimulation (SCS), and Peripheral Nerve Stimulation (PNS) Therapies to include the following.
1. Enabling cycling at certain parameter settings may decrease the recharge interval for rechargeable devices and the device longevity of non-rechargeable devices. Longevity calculations for non-rechargeable devices during SCS trailling with clycling enables as calculated on the N'Vision Clinician Programmer may not be accurate.
2. The current labelling does not address the potential for epidural mass formation and subsequent spinal cord compression.
Recall action type:Alert
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:12/3/2014
Hide menus