Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16353 |
| Brand Name: | Eschman RX Powered Operating Tables |
| Model: | All RX series Operating Tables |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Eschmann Holdings Ltd t/a Eschmann Equipment |
| Issue: | Following a reported incident it has been identified that handsets are being disassembled by users, presumably to facilitate a repair. The disassembly of the handset for any reason can compromise the seal that protects the internal circuits from fluid ingress. Should fluid enter the handset; this can cause an unrequested movement of the operating table top. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/3/2014 |