Detail of Recall

Type of Product:Device
Medsafe Reference:16350
Brand Name:Bio-Rad D-10 Hemoglobin Testing System
Model:Catalogue Number: 220-0220
Affected:
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:
Issue:On a rare occasion the D-10 software might generate an extra result by duplicating the result from the previous sample and the Sample ID/Injection # from the next sample.
The duplicate is readily detectable when sorting the Daily Summary by the injection # and reviewing for duplication of the same sample ID number. We recommend carefully reviewing the Daily Summary and/or the printed chromatograms for duplicate injection numbers or Sample IDs. Such duplication could potentially lead to discrepant sample results, which should not be reported.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/3/2014