Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16350 |
| Brand Name: | Bio-Rad D-10 Hemoglobin Testing System |
| Model: | Catalogue Number: 220-0220 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | |
| Issue: | On a rare occasion the D-10 software might generate an extra result by duplicating the result from the previous sample and the Sample ID/Injection # from the next sample. The duplicate is readily detectable when sorting the Daily Summary by the injection # and reviewing for duplication of the same sample ID number. We recommend carefully reviewing the Daily Summary and/or the printed chromatograms for duplicate injection numbers or Sample IDs. Such duplication could potentially lead to discrepant sample results, which should not be reported. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/3/2014 |