Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16347 |
| Brand Name: | DePuy LCS Complete 3-Peg Metal Backed Patella |
| Model: | Product Number: 129409740 & 129409750 |
| Affected: | Lots: 1858741 & 1858741 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (NZ) Limited 13a Gabador Place Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | DePuy Orthopaedic Technology Inc |
| Issue: | During routine inspection a dimensional discrepancy was observed that alters the mechanical lock between the polyethylene and the biteplate. Due to this, we are advised that such implants had a higher risk of breakage of the polyethylene component or dissociation of the polyethylene articular surface from the baseplate. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 5/3/2014 |