Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16341 |
| Brand Name: | Elekta XiO Radiation Treatement Planning System |
| Model: | |
| Affected: | |
| Software version: | Versions 4.1 to 5.00.00 |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs. The problem occurs because only the start and stop angles of the Dynamic Conformal arcs are computed when all sub-beams of the arc should be computed. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/3/2014 |