Type of Product:Device
Medsafe Reference:16341
Brand Name:Elekta XiO Radiation Treatement Planning System
Model:
Affected:
Software version:Versions 4.1 to 5.00.00
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs. The problem occurs because only the start and stop angles of the Dynamic Conformal arcs are computed when all sub-beams of the arc should be computed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2014