Detail of Recall

Type of Product:Device
Medsafe Reference:16288
Brand Name:Philips HeartStart MRx Monitor/Defib with Q-CPR Meter
Model:Model numbers M3535A and M3536A with Q-CPR™ Meter Option B08
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips Heartstart MRx Monitor/Defib with Q-CPR Meter
Model:Model Numbers: M3535A and M3536A with Q-CPR™ Meter Option B08
Affected:
Software version:
Recalling Organisation:Laerdal (New Zealand) Ltd
79 Apollo Drive
Rosedale
Auckland 0632
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When the HeartStart MRx is in AED mode, the Q-CPR meter is designed to display the 'Do Not Touch the Patient' icon during the shock advisory analysis period. If the user continues to deliver chest compressions when the shock advisory analysis begins, the 'Do Not Touch the Patient' icon may incorrectly remain on the screen after analysis or shock delivery is complete.
When the HeartStart MRx is in Manual Mode, the Q-CPR meter is designed to display the 'Do Not Touch the Patient' icon when the device is charging to the selected energy. If the clinician continues to provide chest compressions while the device is charging, the Q-CPR meter 'Do Not Touch the Patient" icon may incorrectly remain on the screen after shock delivery is complete.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/2/2014