Detail of Recall

Type of Product:Device
Medsafe Reference:16265
Brand Name:Biomerieux Etest Ertapenem foam and Single Blister Packs
Model:Reference Numbers: 531618, 531640, 415332
Affected:Lots: 1002054430, 1000853690 & 1001960650, 1002110560 & 1002389880
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:bioMérieux has identified that the affected Etest strips do not give the correct Minimum Inhibitory Concentration (MIC) for some Quality Control strains, with the MIC results out of range (high) according to the MIC specification described in the Quality Control procedure.
The MIC results conformed at the time of batch release however investigations have confirmed this issue appears over time, leading to a performance issue by overestimating the MIC.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/2/2014