Detail of Recall

Type of Product:Device
Medsafe Reference:16251
Brand Name:Abbott ARCHITECT Intact PTH Calibrators, Controls, and Reagents
Model:Product Codes: 8K25-20, 8K25-25
Affected:Specific lots
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott GmbH & Co. KG
Issue:Falsely elevated Intact PTH results may be generated.
Results generated with affected lots may demonstrate a positive shift relative to that generated with previous reagent and/or calibrator lots. Abbott ARCHITECT Intact PTH Controls do not detect the shift.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/2/2014