Detail of Recall

Type of Product:Device
Medsafe Reference:16240
Brand Name:Raystation RaySearch therapy treatment planning system
Model:
Affected:
Software version:Versions: 2.0, 2.5, 3.0, 3.5, 4.0
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The manufacturer advises there is an issue with the dose calculation in specific versions of the RayStation software. The error appears in some situations for bolus/external/fixation/support structures that are outside the image stack but covered by the dose grid. When converted to dose grid structures they are cut off at the image stack boundaries (first and last slice) in the inferior-superior direction. For a bolus/external/fixation/support structure this means that the part outside the image stack does not contribute to the density distribution and the dose calculation will be affected. Only structures before the first slice or after the last slice are affected, not structures outside the image stack in the anterior/posterior/left/right direction.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/2/2014