Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16238
Brand Name:Raystation RaySearch treatment planning system
Model:
Affected:
Software version:Versions: 2.5, 3.0, 3.5 and 4.0
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The rotation component of the rigid registration computed prior to the deformable registration is not accounted for in fusion views when mapping structures (regions or points of interest) or when deforming dose. Since the error affects all applications of the hybrid deformation, its effect on the above functions is displayed in the user interface. At that time the user would be able to view the result and evaluate the quality of the deformation and identify any potential errors in the results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/2/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /