Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16238 |
| Brand Name: | Raystation RaySearch treatment planning system |
| Model: | |
| Affected: | |
| Software version: | Versions: 2.5, 3.0, 3.5 and 4.0 |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The rotation component of the rigid registration computed prior to the deformable registration is not accounted for in fusion views when mapping structures (regions or points of interest) or when deforming dose. Since the error affects all applications of the hybrid deformation, its effect on the above functions is displayed in the user interface. At that time the user would be able to view the result and evaluate the quality of the deformation and identify any potential errors in the results. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/2/2014 |