Detail of Recall

Type of Product:Device
Medsafe Reference:16221
Brand Name:Siemens 550TxT Treatment Table for Primus, Oncor and Artiste Digital Linear Accelerators
Model:Catalogue Number: 7346534
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:It is possible for patients or operators to have their fingers trapped and pinched between the table top and the bottom frame of the 550 TxT™ Treatment Table.
Secondly and unrelated we have received a report on an incident in which a cable inside of the 550 TxT™ Treatment Table has been damaged due to contacting a moving part. The resulting short-circuit caused overheating of an electronic circuit board inside of the table base.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/2/2014