Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16207 |
| Brand Name: | The Binding Site SPAPLUS Analyser |
| Model: | Catalogue Number: TBSIE610 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | In Vitro Technologies 26-30 Vestey Drive Mount Wellington Panmure Auckland |
| Contact Information: | 0800 468 487 |
| Manufacturer: | The Binding Site Group Ltd |
| Issue: | The manufacturer advises it has received reports of the "Prog Error (overflow) DET Comm" error message on SPAPLUS analysers with the 'C' suffix in the serial number. When the error occurs, analysis will stop and a full shutdown is required to restart testing. In addition to this, it causes results to be lost and the sample/reagents to be wasted. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/2/2014 |