Type of Product:Device
Medsafe Reference:16207
Brand Name:The Binding Site SPAPLUS Analyser
Model:Catalogue Number: TBSIE610
Affected:
Software version:
Recalling Organisation:In Vitro Technologies
26-30 Vestey Drive
Mount Wellington
Panmure
Auckland
Contact Information:0800 468 487
Manufacturer:The Binding Site Group Ltd
Issue:The manufacturer advises it has received reports of the "Prog Error (overflow) DET Comm" error message on SPAPLUS analysers with the 'C' suffix in the serial number. When the error occurs, analysis will stop and a full shutdown is required to restart testing. In addition to this, it causes results to be lost and the sample/reagents to be wasted.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/2/2014