Type of Product:Device
Medsafe Reference:16182
Brand Name:GE Infinia Nuclear Medicine Systems
Model:Models: Infinia, Infinia Hawkeye, Millennium VG, VG Hawkeye, VariCam
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Functional Imaging Haifa
Issue:A potential life-threatening hazardous situation may occur if there is unintended radial detector motion leading to patient contact. This failure mode has been observed on a single system for a Whole Body scan, in which the detector moved to the home position and then slowly started drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. No injuries have been reported to date.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/2/2014