Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16182 |
| Brand Name: | GE Infinia Nuclear Medicine Systems |
| Model: | Models: Infinia, Infinia Hawkeye, Millennium VG, VG Hawkeye, VariCam |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Functional Imaging Haifa |
| Issue: | A potential life-threatening hazardous situation may occur if there is unintended radial detector motion leading to patient contact. This failure mode has been observed on a single system for a Whole Body scan, in which the detector moved to the home position and then slowly started drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. No injuries have been reported to date. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/2/2014 |