Detail of Recall

Type of Product:Device
Medsafe Reference:16159
Brand Name:Philips Heartstart XL+ Defibrillator/ Monitor
Model:861290
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:We have discovered through customer feedback and internal testing that the HeartStart XL+ battery charge time to 100% capacity at 35°C (95°F) is slightly longer than the specified 3 hour duration stated in the existing product labelling. Since the XL+ can operate with a partially charged battery, there is no patient risk associated with a longer battery charge time.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/2/2014