Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16080
Brand Name:Medtronic Activa PC 37601 Neurostimulator
Model:Model 37601
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa SC 37602 Neurostimulator
Model:Model 37602
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa SC 37603 Neurostimulators
Model:Model 37603
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic Activa RC 37612 Neurostimulator
Model:Model 37612
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreULTRA 37712 Neurostimulator
Model:Model 37712
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic RestoreSensor 37714 Neurostimulators
Model:Models 37714
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:1) Unexpected Loss of stimulation may occur under the following conditions: - Switching from a group with two programs to a group with three or four programs where a non-negative contact is shared within the programmed groups. - Creating Program Groups (Activa SC only) when a second program is created for the first time. 2) Under a specific set of conditions (Activa SC - Models 37602 / 37603 are not affected), typically related to device recovery from an over discharge, there is a potential for over stimulation or stimulation directed to a lead electrode other than what was intended: - Device reaches a power on reset (POR) state. A POR may be detected during interrogation with the patient or physician programmer. The patient will experience loss of therapy if a POR occurs. - 'Therapy OFF' command is sent (by patient programmer, INSR, or 'THERAPY-STOP' button on the 8840 clinician programmer) to device while the device is making an automatic periodic battery measurement.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:30/1/2014