Detail of Recall

Type of Product:Device
Medsafe Reference:16069
Brand Name:Sorin Perceval S Post-Dilation Catheter, Size M
Model:Item: ICV1149
Affected:15 lots
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:Sorin Group Italia Srl
Issue:In specific models and lots of this device, the blue marker (placed on the catheter shaft close to the balloon proximal extremity, aiding correct position of the balloon with respect to the prosthesis, prior to dilation) may detach and displace during removal of the plastic tube which protects the balloon. Removal of the plastic tube protection may displace the blue marker distal to the balloon.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/1/2014