Detail of Recall

Type of Product:Device
Medsafe Reference:16064
Brand Name:Covidien Puritan Bennett 840 and 700 Series Ventilators
Model:840201311
Affected:
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Covidien LLC
Issue:Manufacturer conducting a recall for product correction with updated labeling to clarify the operational life of the oxygen sensor (O2 sensor) P/N 4-072214-00 used in the Puritan Bennett 840 ventilator, and P/N G-062010-00 used in Puritan Bennett 700 series ventilators (Puritan Bennett 740 and 760 ventilators).
The oxygen sensor for the ventilators has a life of one (1) year from date of manufacture of the sensor, depending on the operating conditions.
Conflicting information was provided to Puritan Bennett 840 ventilator customers in a labeling addendum stating that the oxygen sensor should be replaced every two years. The correct operational life is one year from the date of manufacture as stated in the operators manual.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/1/2014