Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16063 |
| Brand Name: | Philips BrightView SPECT, BrightView X and BrightView XCT Imaging Systems |
| Model: | 882480 882478 882482 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Radiation Oncology Systems |
| Issue: | Manufacturer has identified several software issues which will be correcetd with the release of a new software version. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/1/2014 |





