Type of Product:Device
Medsafe Reference:16052
Brand Name:Baxter Healthcare Bladder Irrigation Connector
Model:EMC3207
Affected:
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare Corporation SA
Issue:Manufacturer has been reviewing the stability data generated to support labelling claims on its products and found that the microbial barrier requirements, provided by means of the packaging material to guarantee sterility of the product were not met for the named product code. As a result the manufacturer cannot guarantee sterility of the product over the whole claimed shelf life. The medical risk has been assessed as high and therefore manufacturer has made the decision to recall all batches of this product code.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/1/2014