Detail of Recall

Type of Product:Device
Medsafe Reference:16028
Brand Name:Siemens Artiste syngo RT Therapist (RTT) Software
Model:
Affected:
Software version:Version 4.3
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:There is a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist 4.3 software. This feature allows for one or more partially delivered beams to be re-delivered to make up for the deficit in total Monitor Units (MU) planned for delivery.
In a rare scenario in which several conditions are fulfilled, the system will send the already treated beam again to the control console of the linear accelerator for full delivery. This will result in a greater dose being delivered than was planned.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/1/2014