Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 16013 |
| Brand Name: | GE Innova Cardiovascular X-ray System |
| Model: | Models: 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ, 3131IQ |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems SCS |
| Issue: | The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional imaging. No injuries have been reported due to this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/1/2014 |