Detail of Recall

Type of Product:Device
Medsafe Reference:16011
Brand Name:Medtronic Mosaic Heart Valve Sizer Sets
Model:Model Numbers: 7305, 7305OD, 7305UX
Affected:All lots
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:Medtronic has received infrequent reports of higher-than-expected transvalvular gradients occurring post implant.
Medtronic has determined that the practice of substantial oversizing can lead to cases of higher-than-expected transvalvular gradients. Specifically, considerable oversizing, or implanting a valve substantially larger than the native aortic annulus, may result in the alteration of normal leaflet movement as blood flowing through the native annulus may not properly match the size and/or shape of the inflow of the Mosaic Aortic Bioprosthesis.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/1/2014