| Type of Product: | Device |
| Medsafe Reference: | 16008 |
| Brand Name: | GE PET Discovery 600 and Discovery 690 Nuclear Medicine System |
| Model: | 600 & 690 |
| Affected: | |
| Software version: | Software versions pet_mict_plus.44 and pet_velocity.53 |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | A software issue exists when using List Mode Replay, such that subsequent reconstructions with that data could contribute to an image with scatter over-estimation and overcorrection. Reconstructed images with this issue will have many fewer counts than the acquired scan data and a high scatter fraction. This issue may affect studies that use List Mode Replay and are reconstructed after subsequent acquisitions. This issue may lead to image misinterpretation and/or inaccurate treatment decisions |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/1/2014 |