Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:16001
Brand Name:S&N Journey Bi-Cruciate Stabilised femoral implant
Model:All models
Affected:All lots
Software version:
Recalling Organisation:Smith & Nephew Surgical Ltd
Unit A 36 Hillside road
Wairau Valley
AUCKLAND 0760
Contact Information:0800 657 799
Manufacturer:Smith & Nephew Inc
Issue:The Femoral Implant/Tibial Baseplate combination of the JOURNEY BCS Knee System has been identified in the Australian Orthopaedic Association's National Joint Replacement Registry (AOANJRR) as having a higher than expected revision rate when compared to all other primary total knee replacements.. The AOANJRR data indicates that Journey BCS has a higher incidence of revisions due to patellofemoral pain, unspecified pain and instability. The most recent JOURNEY BCS Knee System revision data published by the NJRR include:
- A revision rate of 1.59 (1.34, 1.88) revisions per 100 observed component years compared to 0.72 (0.70, 0.73) for other total knee replacements; and
- A yearly cumulative percent revision of 7.0% (5.8, 8.4) at 5 years compared to 3.8% (3.7, 3.8) for other total knee replacements.
Smith & Nephew is withdrawing the Femoral Implant component of the JOURNEY BCS Knee System from the market.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:10/3/2014