Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15975
Brand Name:Siemens Immulite Androstenedione
Model:Catalogue Numbers: LKAO1, L2KAO2
Affected:Lot numbers: 409 & above, 314 & above (respectively)
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Products Limited
Issue:Siemens Healthcare Diagnostics has confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL (19.2 nmol/L) with the IMMULITE/IMMULITE 1000 and IMMULITE 2000/IMMULITE 2000 XPi Androstenedione assays (LKAO1, L2KAO2).
Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL (19.2 nmol/L), instead of 10 ng/mL (34.9 nmol/L) as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5 ng/mL (19.2 nmol/L) and 10 ng/mL (34.9 nmol/L) will detect this issue.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/1/2014
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /