Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15934 |
| Brand Name: | Siemens Artis Systems with large displays |
| Model: | All Artis systems |
| Affected: | |
| Software version: | VC21B |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | If the system is not turned "OFF" manually at regular intervals an automatic reset of the Large Display Controller can occur. During the reset the system switches for approximately 1-2 minutes into the "Bypass Fluoro" mode. During this time only the native fluoroscopic image is displayed. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/12/2013 |