Detail of Recall

Type of Product:Device
Medsafe Reference:15934
Brand Name:Siemens Artis Systems with large displays
Model:All Artis systems
Affected:
Software version:VC21B
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:If the system is not turned "OFF" manually at regular intervals an automatic reset of the Large Display Controller can occur. During the reset the system switches for approximately 1-2 minutes into the "Bypass Fluoro" mode. During this time only the native fluoroscopic image is displayed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2013