Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15932
Brand Name:Elekta Integrity
Model:
Affected:
Software version:R1.1, R3.0, and R3.1
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values. If you do not do the quality assurance (QA) procedures after the calibration, it is possible that you will not find the error made during the calibration
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/12/2013
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /