Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15897 |
| Brand Name: | GE Giraffe Infant Warmers with blenders |
| Model: | All models with blender |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Ohmeda Medical Inc |
| Brand Name: | GE Panda Infant Warmers with blenders |
| Model: | All units with blender |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Ohmeda Medical Inc |
| Issue: | The manufacturer has become aware of a potential safety issue associated with the loosening of the blender knob shaft on all Resuscitation units. A loose blender knob shaft may potentially affect the accuracy of the blended output of Air/O2 gas mixture from the unit. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/12/2013 |