Detail of Recall

Type of Product:Device
Medsafe Reference:15885
Brand Name:HeartWare Ventricular Assist System
Model:1100, 1101, 1102,1103, 1104 and 1205
Affected:
Software version:
Recalling Organisation:Health New Zealand - Te Whatu Ora
9 Ruskin Street
Parnell
Auckland 1052
Contact Information:
Manufacturer:HeartWare International Inc
Issue:Manufacturer has discovered a failure mechanism arising from improper thread locking technique. In 8 cases customers have reported that the locking mechansim of the driveline connector has failed to engage . Of these 8 reports none has resuklted in any injury.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:18/12/2013