Detail of Recall

Type of Product:Device
Medsafe Reference:15878
Brand Name:Philips Essenta DR Digital X-Ray, C-arm
Model:All Essenta DR systems
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:If during system installation, the screws holding the C-arm were mounted using thread locking compound (e.g. LOCTITE), then the corresponding thread inserts inside the C-arm casting may loosen when the C-arm is dismantled. If, after reassembly of the system, the loosened thread inserts become detached from the casting, the C-arm would no longer securely held in position and could fall. The C-arm is most often dismantled and reassembled when the system is moved to another location.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/12/2013