Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15868 |
| Brand Name: | GE Optima XR220amx & XR220amx with Digital Upgrade X-Ray Systems |
| Model: | 5555000-5, 5555000-6 and 5421698 |
| Affected: | |
| Software version: | All versions |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient. If the user does not notice this change and continues to perform the exam, the images will be placed in an incorrect patient file. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/12/2013 |