Detail of Recall

Type of Product:Device
Medsafe Reference:15868
Brand Name:GE Optima XR220amx & XR220amx with Digital Upgrade X-Ray Systems
Model:5555000-5, 5555000-6 and 5421698
Affected:
Software version:All versions
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient. If the user does not notice this change and continues to perform the exam, the images will be placed in an incorrect patient file.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/12/2013