Detail of Recall

Type of Product:Device
Medsafe Reference:15857
Brand Name:Siemens Somatom Computed Tomography System
Model:Definition, Flash, Edge, AS
Affected:
Software version:Syngo version CT2012B or CT2013B
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens syngo CT Software
Model:
Affected:
Software version:Syngo version CT2012B and CT2013B
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:If the function "Auto save last displayed images as key images" is activated during an interventional procedure, the software configuration allows conflicting access to the CT system image database. In such situations a system display freeze might occur. Normal interaction with the system is then no longer possible. No error message is issued and only a system restart resolves the problem. In case of a sudden system display freeze the interventionalist will potentially not immediately recognize the malfunction and continue the interventional procedure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/12/2013