Detail of Recall

Type of Product:Device
Medsafe Reference:15856
Brand Name:Siemens RAPIDLyte Arterial Blood Sampling Line Draw Syringes
Model:Catalogue Numbers: 04180028, 00355281, 06194484, 06380806
Affected:Various lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Smiths Medical ASD Inc
Issue:The manufacturer has identied an issue with the ABL syringes when used with Radiometer brand blood gas analysers. The plunger tip in the ABL syringe does not remain stationary when the probe of the analyzer extends into the syringe. This can result in difficulties aspirating the sample or the plunger may get pushed out of the syringe.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/12/2013